CLEANLINE fittings for 5.0 gases
- general information

BOSS CLEANLINE 5.0 fittings can be used for gases up to 5.0 (99,999% clean gas).

The fittings are used when the requirements for gas purity are high: Laboratories in general, chromatography, Ar, He, N2, H2, CH4 and working gases for spectrometry.

BOSS CLEANLINE 5.0 special features:
  • Ultrasonically cleaned components
  • Sealings are made of PVDF and FPM (Viton)
  • Packaged in sealed bags to maintain cleanliness
  • Twin ferrule compression fittings secure clean connection on outlet.

All fittings are leak tested before leaving the factory. Leak rate better than 10-3 mbarl/s.
1 out of 100 valves is helium leak tested. Leak rate better than 5x10-5 mbar l/s.
Maximum test pressure 16 bar / 235 psi.

BOSS CLEANLINE 5.0 Stop valve for shut-off of media flow.
Opening / closing function: 1.5 x 360º turn.



Materials with medium contact:
Metals: Brass
Sealing: Viton, PVDF
Lubricant: Approved for oxygen

Capacity graph

BOSS CLEANLINE 5.0 pressure regulation valve for regulation of media pressure. With built-in pressure gauge.
The pressure regulator is available as 3 bar and 8 bar versions.

Materials with medium contact:
Metals: Brass and stainless steel
Sealing: EPDM, PA

Flow chart

BOSS CLEANLINE 5.0 flow regulation valve. For regulation of media flow.
Opening / closing function: 3 x 360º turn..



Materials with medium contact:
Metals: Brass
Sealing: Viton, PVDF
Lubricant: Approved for oxygen

Capacity graph